Remote Commands Reference
R&S
®
RTC1000
244
User Manual 1335.7352.02 ─ 02
MASK:CHCopy
Creates a mask from the envelope waveform of the test source set with
.
Usage:
Event
MASK:YPOSition
<Yposition>
Moves the mask vertically within the display.
Parameters:
<Yposition>
Mask offset from the vertical center
Range:
-200 to 200
*RST:
0
Default unit: div
MASK:YSCale
<Yscale>
Changes the vertical scaling to stretch or compress the mask in y-direction.
Parameters:
<Yscale>
A value over 100
% stretches the amplitudes; a value less than
100
% compresses the amplitudes.
Range:
10 to 1000
*RST:
100
Default unit: %
MASK:YWIDth
<Yaddition>
Changes the width of the mask in vertical direction.
Parameters:
<Yaddition>
The value is added to the y-values of the upper mask limit and
subtracted from the y-values of the lower mask limit.
Range:
0 to 5.12
*RST:
0
Default unit: div
MASK:XWIDth
<Xaddition>
Changes the width of the mask in horizontal direction.
Masks