Instrument tour
R&S
®
RTE
34
Getting Started 1326.1026.02 ─ 10
Chapter 4.7, "Switching on or off"
5.3.2
SETUP controls
SETUP keys set the instrument to a defined state, change basic settings, and
provide saving and help functions. The intensity rotary knob adjusts the display
contrast for several display elements.
[AUTOSET]
The instrument analyzes the enabled channel signals, and adjusts appropriate
horizontal, vertical, and trigger settings to display stable waveforms.
[PRESET]
Resets the instrument to a default state. All measurements, mask tests, zoom,
and most individual settings are deleted, and all channels except for channel 1
are disabled. You can define preset configurations and save them to a file. The
[PRESET] key can be configured to set either factory defaults or a user-defined
preset configuration.
[FILE]
Opens and closes the "File" dialog box, where you can:
●
Save instrument settings (user settings)
●
Load instrument settings which were saved before
●
Save waveform data and measurement results
●
Define a naming pattern for autonaming of files
[SETUP]
Opens and closes the "Setup" dialog box, where you can:
●
Access Windows configuration and install firmware updates
●
Configure the touchscreen
●
Check and install option keys for software options
●
Check availability of hardware options
●
Configure remote settings, LAN settings, and GPIB
[PRINT]
Starts a printing or saving action, or opens a report. The function is assigned to
the key in "File" menu > "Frontpanel Setup" > "Hardkeys". By default, the key
saves a screenshot of the waveform display.
Keys and controls