Trying out the instrument
R&S
®
RTE
47
Getting Started 1326.1026.02 ─ 10
8. Tap the "Undo" icon on the toolbar repeatedly until the trigger mode is "Auto"
and the trigger source is "CH1".
9. Press the [SLOPE] key to toggle the trigger slope.
Watch the waveform and the "Trigger" settings.
6.2
Acquiring data
You can acquire data using different arithmetic methods to get envelope or aver-
age waveforms.
If you need to change only one setting during analysis, and you need to change it
often, you can reduce the dialog box to a small box that only contains the
required setting. Thus you can change the setting and see the result immediately.
1. Press the [ACQUISITION] key on the front panel, in the [HORIZONTAL] area.
2. In the "Acquisition" tab of the "Horizontal" dialog box, select the "Decimation"
type
Sample
.
3. Tap the "Arithmetic" button for C1 but do not change the setting.
4. Tap the "Minimize" icon in the upper right corner of the dialog box.
Acquiring data