Instrument tour
R&S
®
RTE
37
Getting Started 1326.1026.02 ─ 10
"Horizontal position" defines the time distance of the reference point from the zero
point of the diagram. Turn clockwise to move the waveform to the right.
"Reference point" defines the position of the reference point on the screen. Turn
clockwise to move it to the right. The reference point marks the rescaling center
of the time scale. It is indicated by a gray triangle outline at the top of the dia-
gram. If you modify the time scale, the reference point remains fixed on the
screen, and the scale is stretched or compressed to both sides of the reference
point.
[SCALE]
The rotary knob adjusts the time scale for all signals. The time scale is also
known as timebase.
Turn clockwise to stretch the waveforms. Doing so, the scale value
time/div
decreases.
Press the knob to toggle between coarse and fine scale adjustment.
5.3.4
VERTICAL controls
The keys and knobs in the VERTICAL functional block select a signal and adjust
the vertical scale and position of the selected signal.
[CH ×]
Turns on, selects, and configures a channel. If the channel is active, the key lights
up in the corresponding channel color .
The effect of the keypress depends on state of the channel:
●
If channel is off: Pressing the key turns on the channel and selects it.
●
If the channel is on, but not selected: Pressing the key selects the channel
waveform.
Keys and controls