Remote Control Commands
R&S
®
SGT100A
332
User Manual 1176.8674.02 ─ 07
Parameters:
<ApTime>
float
Range:
depends on connected power sensor
Increment: 1E-9
*RST:
depends on connected power sensor
Example:
SENS:POW:APER:TIM 23ms
Sets 23 ms aperture time.
Manual operation:
See
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe
<State>
Activates the use of the S-parameter correction data.
Note:
If you use power sensors with attenuator, the instrument automatically activates
the use of S-parameter data.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
SENS1:POW:CORR:SPD:STAT ON
Activates the use of the S-parameters correction data of power
sensor 1.
Manual operation:
See
:SENSe<ch>[:POWer]:DISPlay:PERManent:PRIority
<Priority>
Selects average or peak power for permanent display.
Parameters:
<Priority>
AVERage | PEAK
*RST:
AVERage
Example:
SENS1:DISP:PERM:STAT ON
The permanent viewer is switched on.
SENS1:DISP:PERM:PRI AVER
The measured average power is indicated.
Manual operation:
See
:SENSe<ch>[:POWer]:DISPlay:PERManent:STATe
<State>
Activates the permanent display of the measured power level results. The instrument
also indicates the sensor type, the connection, the measurement source and the offset
if set.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0