Remote Control Commands
R&S
®
SGT100A
361
User Manual 1176.8674.02 ─ 07
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:TRIGger[:EXTernal<1|2>]:INHibit
<Inhibit>
Specifies the number of samples by which a restart is to be inhibited following a trigger
event. This command applies only in the case of external triggering.
Parameters:
<Inhibit>
integer
Range:
0 to 67108863
*RST:
0
Default unit: sample
Example:
BB:ARB:TRIG:SOUR EXT
selects an external trigger
BB:ARB:TRIG:INH 200
sets a restart inhibit for 200 samples following a trigger event.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary[:TRIGger]:SEQuence
<Sequence>
The command selects the trigger mode.
SOURce Subsystem