Remote Control Commands
R&S
®
SGT100A
311
User Manual 1176.8674.02 ─ 07
●
Activating the original state of delivery (factory reset,
on page 311). Only functions that are protected by a password remain unchanged
and the passwords themselves.
:SOURce<hw>:PRESet
:SYSTem:PRESet
Triggers an instrument reset. It has the same effect as:
●
The
*RST
command
Example:
SYST:PRES
All instrument settings (also the settings that are not currently
active) are reset to their default values.
Usage:
Setting only
:SYSTem:FPReset
Triggers an instrument reset to the original state of delivery.
Example:
SYST:FPR
all instrument settings (also those that are not currently active)
are reset to the factory values.
Usage:
Event
Manual operation:
See
11.4 CALibration Subsystem
:CALibration<hw>:BBIN[:MEASure]?
................................................................................311
......................................................................................... 312
:CALibration:FREQuency[:MEASure]?
..............................................................................312
:CALibration:IQModulator:FULL?
..................................................................................... 312
:CALibration:IQModulator:LOCal?
....................................................................................312
.......................................................................................313
:CALibration:FREQuency:TEMPerature?
..........................................................................313
:CALibration:LEVel:TEMPerature?
...................................................................................313
:CALibration:IQModulator:TEMPerature?
..........................................................................313
:CALibration<hw>:ROSCillator[:DATA]
............................................................................. 313
:CALibration<hw>:BBIN[:MEASure]?
Starts adjustment of the analog I/Q input. The I/Q input is adjusted with respect to DC
offset and gain.
Return values:
<Measure>
0 | 1 | OFF | ON
*RST:
0
CALibration Subsystem