Varying the RF Signal in List or Sweep Mode
R&S
®
SMA100B
113
User Manual 1178.3834.02 ─ 03
●
Switches automatically to the next step when the
has elapsed.
●
Stops signal generation at the set end value and waits for the subsequent trigger
event.
●
Trigger sources:
–
The "Execute Single" function.
–
The corresponding remote control command.
–
An externally applied trigger signal.
●
"State = Off" stops the signal generation in sweep or list mode.
Table 5-2: Cross-reference between manual and remote control in Single / Extern Single modes
(Sweep/List)
Manual control mode:
"Single / Extern Single"
Remote commands
Rohde
&
Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:TRIGger<hw>:FSWeep:SOURce SINGle [BUS]
"Single"
or
:TRIGger<hw>:FSWeep:SOURce EXTernal [EXTernal]
"Extern Single"
:SOURce<hw>:SWEep:FREQuency:MODE AUTO
RF level sweep
:TRIGger<hw>:PSWeep:SOURce SINGle [BUS]
for "Single"
or
:TRIGger<hw>:PSWeep:SOURce EXTernal [EXTernal]
for "Extern Sin-
gle"
:SOURce<hw>:SWEep:POWer:MODE AUTO
LF frequency sweep
:TRIGger<hw>:LFFSweep:SOURce SINGle [BUS]
"Single"
or
:TRIGger<hw>:LFFSweep:SOURce EXTernal [EXTernal]
"Extern Sin-
gle"
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE AUTO
List
:SOURce<hw>:LIST:TRIGger:SOURce SINGle
"Single"
or
:SOURce<hw>:LIST:TRIGger:SOURce EXTernal
"Extern Single"
:SOURce<hw>:LIST:MODE AUTO
Signal Generation and Triggering in the Sweep and List Modes