Varying the RF Signal in List or Sweep Mode
R&S
®
SMA100B
115
User Manual 1178.3834.02 ─ 03
Table 5-3: Cross-reference between manual and remote control in Step / Extern Step modes (Sweep/
List)
Manual control mode:
"Step / Extern Step"
Remote commands
Rohde
&
Schwarz proprietary and [SCPI conform] value name
RF frequency sweep
:TRIGger<hw>:FSWeep:SOURce SINGle [BUS]
"Step"
or
:TRIGger<hw>:FSWeep:SOURce EXTernal [EXTernal]
for "Extern
Step"
:SOURce<hw>:SWEep:FREQuency:MODE STEP
RF level sweep
:TRIGger<hw>:PSWeep:SOURce SINGle [BUS]
"Step"
or
:TRIGger<hw>:PSWeep:SOURce EXTernal [EXTenal]
"Extern Step"
:SOURce<hw>:SWEep:POWer:MODE STEP
LF frequency sweep
:TRIGger<hw>:LFFSweep:SOURce SINGle [BUS]
"Step"
or
:TRIGger<hw>:LFFSweep:SOURce EXTernal [EXTenal]
"Extern Step"
:SOURce<hw>:LFOutput<ch>:SWEep:FREQuency:MODE STEP
List
:SOURce<hw>:LIST:TRIGger:SOURce SINGle
"Step"
or
:SOURce<hw>:LIST:TRIGger:SOURce EXTernal
"Extern Step"
:SOURce<hw>:LIST:MODE STEP
Extern Start/Stop mode (sweep)
Figure 5-6: Extern Start/Stop mode (sweep)
●
The instrument generates the signal continuously.
●
Trigger mode "Auto" and "Sweep > State = On" are prerequisite.
●
State "On" sets the signal to the start value, that is one of the following:
–
Sweep start frequency
–
Sweep start power
–
Frequency-power value pair of the selected index in the list
●
Starts signal generation with a trigger event.
Signal Generation and Triggering in the Sweep and List Modes