Varying the RF Signal in List or Sweep Mode
R&S
®
SMA100B
130
User Manual 1178.3834.02 ─ 03
See
Chapter 5.2.1, "Correlating Parameters in Sweep Mode"
"Step Linear"
The step width is a constant value in Hz.
"Step Logarithmic"
The step width is determined logarithmically in %, i.e. as a constant
fraction of the current frequency.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:STEP[:LINear]
[:SOURce<hw>]:SWEep[:FREQuency]:STEP:LOGarithmic
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:LINear]
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:LOGarithmic
5.5.3 Level Range Settings
Access:
1. Select "Sweep" > "Level Sweep"
2. Select "Level Range".
Settings
.............................................................................................. 130
Start Level / Stop Level
Defines the level sweep range by setting the start and end values.
See
Chapter 5.2.1, "Correlating Parameters in Sweep Mode"
Remote command:
Spacing
Indicates that the level step intervals are linearly (straight proportional) distributed, i.e.
intervals are of equal size.
Set the step size with the parameter "Step".
Sweep Mode Settings