Varying the RF Signal in List or Sweep Mode
R&S
®
SMA100B
138
User Manual 1178.3834.02 ─ 03
5.7 List Editor
The "User Correction" and "List Mode" dialogs provide a build-in list editor for defining
frequency/level value pairs.
The list editors in these two dialogs are similar. The following description shows the
"List Data Editor".
Access:
1. "Sweep" > "List Mode" > "List Mode Data" > "Edit List Mode Data"
2. "Level" > "User Correction" > "Edit User Cor. Data"
List mode data editor
User correction data editor
The editor is a table with RF frequency and power value pairs and standard navi-
gation functions.
The "Edit List Mode Data" dialog has an extra column for defining variable dwell
times.
All values in one row have to be defined. Rows with missing values are ignored and
not saved. Values of incomplete rows get lost.
If you use
global dwell time in list mode
, consider also the following:
●
The instrument uses the value set with
for all list steps.
●
To save the list, however, you must fill in the "Dwell / s " column in each row,
although the values are not used for generating the signal.
Tip:
function to fill the dwell time column automati-
cally.
Settings
..............................................................................................139
............................................................................................... 139
.............................................................................................. 139
List Editor