Improving Level Performance
R&S
®
SMA100B
152
User Manual 1178.3834.02 ─ 03
All values in one row have to be defined. Rows with missing values are ignored and
not saved. Values of incomplete rows get lost.
If you use
global dwell time in list mode
, consider also the following:
●
The instrument uses the value set with
for all list steps.
●
To save the list, however, you must fill in the "Dwell / s " column in each row,
although the values are not used for generating the signal.
Tip:
function to fill the dwell time column automati-
cally.
Settings
..............................................................................................153
............................................................................................... 153
.............................................................................................. 153
Edit List Mode Data
Table with correction or list values.
"Frequency /Hz"
Sets the frequency values.
Remote command:
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency
"Power /dBm"
Sets the level values.
Remote command:
[:SOURce<hw>]:CORRection:CSET:DATA:POWer
"Dwell /s"
In list mode, sets the dwell time values.
Remote command:
Data handling keys
Standard functions for file and data handling.
Note:
Save a list only after filling both columns (frequency and level), otherwise the
entries are lost.
Goto ← Data handling keys
Selects a row for editing.
Edit ← Data handling keys
Enables you to insert, or delete a row or ranges within a list, and provides access to a
dialog for automatic filling.
User Correction