Improving Level Performance
R&S
®
SMA100B
155
User Manual 1178.3834.02 ─ 03
Indicates the used list.
●
"Include Zeroing"
Performs a zeroing procedure before acquiring the user correction data to improve
precision.
No signal may be applied to the sensor during zeroing. RF output is temporarily
switched off during that time.
When unchecked, the zeroing procedure is skipped. However, the RF signal level
might be blanked shortly. This setting is useful if blanking of RF is undesirable or
the absence of power at the sensor cannot be guaranteed.
●
"Execute"
The "Execute" button is only enabled if a sensor is detected and the user correction list
contains at least one frequency value.
Remote command:
[:SOURce<hw>]:CORRection:ZERoing:STATe
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Used SMAB Settings For Measurement
Displays the settings relevant for the measurement.
"RF Source"
Shows the path for which the correction menu settings are made.
"Modulation"
Indicates the modulation state
"Amplitude"
Shows the currently set level.
"Use SParameter"
Indicates whether SParameter correction is used.
"Attenuator Mode Path 1"
Displays the selected mode of the attenuator.
"Level range"
Shows the level range.
Remote command:
n.a.
6.3.4 Import/Export List Files
Access:
1. Select one of the following:
a) "Sweep" > "List mode".
b) "Level" > "Level > User Correction".
c) "Modulation > Pulse Modulation > Pulse Generator > Pulse Mode = Train".
User Correction