Reference Oscillator
R&S
®
SMA100B
180
User Manual 1178.3834.02 ─ 03
.........................................................................................181
Nominal Synchronization Bandwidth
...........................................................................181
.............................................................................................181
................................................................................................ 181
.................................................................................................181
Set to Default
Calls the default settings.
Remote command:
Source
Selects the reference frequency source.
"Internal"
Uses the internal reference oscillator, either with calibrated or a user-
defined adjustment value.
"External"
Uses an external reference signal.
The frequency of the external reference signal must be set with the
parameter "External Reference Frequency".
See
"External Reference Frequency"
Remote command:
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
Deactivate RF Output (if external reference is missing)
Turns off the RF output when the external reference signal is selected, but no signal is
supplied.
This function prevents that no improper RF signal due to the missing external refer-
ence signal is used for measurements. A message indicates that the RF output is
deactivated.
Remote command:
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]
External Reference Frequency
Sets the frequency of the external reference signal.
"Variable"
Option: R&S
SMAB-K704
The external reference signal has an arbitrary frequency, within the
permissible range from 1
MHz to 100
MHz.
Set the output frequency with the paramaters in the
dialog.
"10 MHz, 100 MHz, 1GHz"
Selects the external reference frequency.
Remote command:
[:SOURce]:ROSCillator:EXTernal:FREQuency
Reference Frequency Settings