Reference Oscillator
R&S
®
SMA100B
183
User Manual 1178.3834.02 ─ 03
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
7.5 Adjustments Settings
Access:
1. Select "Frequency" > "Reference Freq"
2. Select "Adjustment".
Settings:
............................................................................................... 183
Adjustment Active
Selects the adjustment mode.
"OFF"
Uses the calibrated internal reference frequency.
"ON"
Allows you to apply a deviation to the internal reference frequency,
according to your requirements.
Enter the value in the
field.
Remote command:
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]
Adjustment DAC Value
Sets a user-defined adjustment value for the internal reference frequency. This value
takes effect when it is activated with
●
"0" represents the calibrated state.
●
The setting range depends on the reference oscillator type and its factory calibra-
tion value.
Note:
The value is not affected by an instrument preset (PRESET key or
*RST
) and the
"Save/Recall" function. It is reset only by factory preset.
Remote command:
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue
Adjustments Settings