Network Operation and Remote Control
R&S
®
SMA100B
275
User Manual 1178.3834.02 ─ 03
5. Use the mouse cursor and keyboard to access the functionality of the instrument
as you would directly perform on the instruments touchscreen and front panel.
11.8.2 Establishing a Remote Control Connection over LAN Using VXI-11
Protocol
In this example, the I/O software library R&S
VISA from Rohde
&
Schwarz is used to
set up a LAN remote control link and remotely control the R&S
SMA100B. R&S
VISA is
running on a controller PC with Windows operating system. When the connection is set
up you can send commands to the instrument and receive the responses.
The remote control connection requires a VISA installation but no additional hardware
on the controller PC. The LAN I/O channel is selected at initialization time using the
VISA resource string (also referred to as "address string"). A VISA alias (short name) is
used to replace the complete resource string. The host address is the R&S
SMA100B's
hostname or its IP address.
See also
Chapter 11.2.1, "LAN Interface"
In this following, we assume that:
●
A LAN remote control link between the controller and the R&S
SMA100B is set up.
●
The R&S
VISA program is installed on the remote PC
http://www.rohde-schwarz.com/rsvisa
Setting up the controller with R&S VISA
To remote control the R&S
SMA100B, we use the R&S
VISA Tester application. The
application communicates via TCP/IP protocol.
The instrument is preconfigured for networks using DHCP (dynamic host configuration
protocol). If this configuration is used, enter the computer name in the position of the IP
address.
1. Connect the controller and the instrument in the same network (network cable) and
switch them on.
2. On the controller, start "R&S
VISA > Tester 32bit" or "R&S
VISA > Tester 64bit".
How to Set Up a Remote Control Connection