Network Operation and Remote Control
R&S
®
SMA100B
283
User Manual 1178.3834.02 ─ 03
b) Reset the instrument
viRST (generator%, "*RST;*CLS")
c) Set 2 GHz frequency
viPrintf (..., "SOUR:FREQ 2GHz\n")
d) Set 20 dBm output level
viPrintf (..., "SOUR:POW -20dBm\n")
The RF frequency and signal level of the instrument are changed.
3. To return to manual operation, press the LOCAL key.
11.9 Tracing SCPI Commands and Messages Exchanged
via the LXI Web Browser Interface
The remote trace functionality allows you to trace commands and messages
exchanged via a remote control interface of the R&S
SMA100B.
To activate the SCPI remote trace:
1. Start a web browser that supports html5 (W3C compliant).
2. Enter the IP address of the R&S
SMA100B in the browser's address bar.
The R&S
SMA100B's welcome page is displayed.
3. In the navigation pane, select "Diagnostics > SCPI Remote Trace".
4. In the toolbar bar of the "SCPI Remote Trace" page, select "live mode > on" and
"logging > on".
"live mode > on" displays all commands and responses, and "logging > on" also
traces messages.
If you now control the R&S
SMA100B with SCPI commands, using an appropriate
tool, the LXI function records the information sent and received.
The function records all sent commands, received responses and messages, and
saves them in an internal database. If "live mode" is disabled, you can display the
recent traces upon request, using the "refresh" button. You can also save the log in
a file.
Note:
The diagnostics functionality will be extended in later releases, e.g. to down-
load or upload SCPI command files from / to the instrument.
11.10 How to Return to Manual Operation
Before returning to manual control, command processing must be completed. Other-
wise, the instrument switches back to remote control immediately.
How to Return to Manual Operation