Remote Control Commands
R&S
®
SMA100B
380
User Manual 1178.3834.02 ─ 03
Manual operation:
See
[:SOURce<hw>]:PULM:IMPedance
<Impedance>
Sets the impedance for the external pulse trigger and pulse modulation input.
Parameters:
<Impedance>
G50 | G10K
*RST:
G50
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:THReshold
<Threshold>
Sets the threshold for the input signal at the PULSE EXT connector.
Parameters:
<Threshold>
float
Range:
0 to 2
Increment: 0.1
*RST:
1
Default unit: V
Example:
SOURce:PULM:THReshold 1
Manual operation:
See
[:SOURce<hw>]:PULM:MODE
<Mode>
Selects the mode for the pulse modulation.
Parameters:
<Mode>
SINGle | DOUBle | PTRain
SINGle
Generates a single pulse.
DOUBle
Generates two pulses within one pulse period.
PTRain
Generates a user-defined pulse train.
Specify the pulse sequence with the commands:
[:SOURce<hw>]:PULM:TRAin:ONTime
[:SOURce<hw>]:PULM:TRAin:OFFTime
[:SOURce<hw>]:PULM:TRAin:REPetition
*RST:
SINGle
Example:
See
Example "Perform pulse modulation"
Options:
PTRain requires R&S
SMAB-K27
Manual operation:
See
SOURce Subsystem