Remote Control Commands
R&S
®
SMA100B
391
User Manual 1178.3834.02 ─ 03
Manual operation:
See
[:SOURce<hw>]:CORRection:CSET:DATA:POWer
<Power>
Enters the level values to the table selected with
.
Parameters:
<Power>
Power#1[, Power#2, ...]
String of values with default unit dB.
Example:
See
Example "Create a table with user-defined correction values
Manual operation:
See
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of frequency/level values in the selected table.
Return values:
<Points>
integer
Range:
0 to 10000
*RST:
0
Example:
See
Example "Create a table with user-defined correction values
Usage:
Query only
[:SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
Fills the selected user correction table with the level values measured by the power
sensor for the given frequencies.
To select the used power sensor set the suffix in key word
SENSe
.
Example:
See
Example "Fill user correction data with sensor"
Usage:
Event
Manual operation:
See
" Fill User Correction Data with Sensor"
[:SOURce<hw>]:CORRection:CSET[:SELect]
<Filename>
Selects or creates a file for the user correction data.
If the file with the selected name does not exist, a new file is created.
SOURce Subsystem