Remote Control Commands
R&S
®
SMA100B
414
User Manual 1178.3834.02 ─ 03
Parameters:
<Mode>
AUTO | MANual | STEP
AUTO
Performs a complete sweep cycle from the start to the end value
when a trigger event occurs.
The dwell time determines the time period until the signal
switches to the next step.
MANual
Performs a single sweep step when a manual trigger event
occurs.
The trigger system is not active. To trigger each frequency step
of the sweep individually, use the command
STEP
Each trigger command triggers one sweep step only.
The frequency increases by the value set with the coammnds:
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP[:
LINear]
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:STEP:
LOGarithmic
*RST:
AUTO
Example:
See
Manual operation:
See
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:POINts
<Points>
Sets the number of steps in an LF sweep.
For information on how the value is calculated and the interdependency with other
parameters, see
Chapter 5.2.1, "Correlating Parameters in Sweep Mode"
Parameters:
<Points>
integer
Range:
2 to POINts
Increment: 1
*RST:
100
Example:
See
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:RETRace
<State>
Activates that the signal changes to the start frequency value while it is waiting for the
next trigger event.
You can enable this feature, when you are working with sawtooth shapes in sweep
mode "Single" or "External Single".
SOURce Subsystem