Remote Control Commands
R&S
®
SMA100B
422
User Manual 1178.3834.02 ─ 03
[:SOURce<hw>]:LIST:INDex:STARt
<Start>
[:SOURce<hw>]:LIST:INDex:STOP
<Stop>
Sets the start and stop index of the index range which defines a subgroup of fre-
quency/level value pairs in the current list.
Parameters:
<Start>/<Stop>
integer
Index range
Only values inside this range are processed in list mode
Range:
0 to list length
*RST:
0
Example:
See
Example "Create an RF list and activate the list mode"
Manual operation:
See
[:SOURce<hw>]:LIST:MODE
<Mode>
Sets the list mode.
The instrument processes the list according to the selected mode and trigger source,
see
LIST:TRIG:SOUR AUTO, SING or EXT
.
Parameters:
<Mode>
AUTO | STEP
AUTO
Each trigger event triggers a complete list cycle.
STEP
Each trigger event triggers only one step in the list processing
cycle. The list is processed in ascending order.
In this mode, you can select between
.
*RST:
AUTO
Example:
See
Example "Create an RF list and activate the list mode"
Manual operation:
See
[:SOURce<hw>]:LIST:POWer
<Power>
Enters the level values in the selected list. The number of level values must corre-
spond to the number of frequency values. Existing data is overwritten.
SOURce Subsystem