Remote Control Commands
R&S
®
SMA100B
424
User Manual 1178.3834.02 ─ 03
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:LIST:TRIGger:SOURce
<Source>
Selects the trigger source for processing lists.
The names of the parameters correspond to those in sweep mode. SCPI standard
uses other names for the parameters; these names are also accepted by the instru-
ment. The SCPI names should be used if compatibility is an important consideration.
For an overview, see the following table:
Rohde
&
Schwarz name
SCPI name
Command in manual control
AUTO
IMMediate
MODE AUTO
SINGle
BUS
MODE SINGLE
or
STEP
EXTernal
EXTernal
MODE EXT TRIG SINGLE
or
EXT TRIG STEP
Parameters:
<Source>
SINGle | BUS | AUTO | EXTernal | IMMediate
AUTO|IMMediate
The trigger is free-running, i.e. the trigger condition is fulfilled
continuously. The selected list is restarted as soon as it is fin-
ished.
SINGle|BUS
The list is triggered by the command
EXTernal
The list is triggered externally and executed once.
*RST:
AUTO
Example:
See
Example "Create an RF list and activate the list mode"
Manual operation:
See
[:SOURce<hw>]:LIST:RUNNing?
Queries the current state of the list mode.
Return values:
<State>
0 | 1 | OFF | ON
1
Signal generation based on the list mode is active.
Example:
See
Example "Create an RF list and activate the list mode"
Usage:
Query only
SOURce Subsystem