[:SOURce<hw>]:POWer:ALC:SEARch
<Search>
Recalculates the instrument internal settings optimized for the current level. Not
required for automatic modes.
Parameters:
<Search>
ON
*RST:
ON
Example:
POW:ALC:SEAR ON
Executes readjustment.
[:SOURce<hw>]:POWer:ALC:SONCe
Activates level control for correction purposes temporarily.
Example:
POW:ALC OFF
Deactivates automatic level control at the RF output.
POW:ALC:SONC
Executes level control (once).
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:POWer:ALC[:STATe]
<State>
Parameters:
<State>
0 | OFF | AUTO | 1 | ON | ONTable | PRESet | OFFTable
AUTO
Adjusts the output level to the operating conditions automati-
cally.
1|ON
Activates internal level control permanently.
OFFTable
Controls the level by means of attenuation values of the internal
ALC table.
0|OFF
Provided only for backward compatibility with other
Rohde
&
Schwarz signal generators.
The R&S
SMA100B accepts these values and maps them auto-
matically as follows:
0|OFF = OFFTable
ONTable
Starts with the attenuation setting from the table and continues
with automatic level control.
*RST:
AUTO
Example:
POW:ALC ON
Activates internal level control.