Remote Control Commands
R&S
®
SMA100B
436
User Manual 1178.3834.02 ─ 03
Manual operation:
See
[:SOURce<hw>]:POWer:ATTenuation:RFOFf:MODE
<Mode>
Selects the state the attenuator is to assume if the RF signal is switched off.
Parameters:
<Mode>
UNCHanged | FATTenuation
FATTenuation
The attenuator switches to maximum attenuation
UNCHanged
Retains the current setting and keeps the output impedance
constant during RF off.
*RST:
n.a. (factory preset: FATTenuation)
Example:
SOURce1:POWer:ATTenuation:RFOFf:MODE FATTenuation
// maximum attenuation is used when the RF output is deactivated
Manual operation:
See
[:SOURce<hw>]:POWer:EMF:STATe
<State>
Displays the signal level as voltage of the EMF. The displayed value represents the
voltage over a 50 Ohm load.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
POW:EMF:STAT 1
Activates voltage level display.
Manual operation:
See
" Display Level as Voltage of EMF "
[:SOURce<hw>]:POWer:LBEHaviour
<Behaviour>
Selects the level behavior at the RF output over time.
Parameters:
<Behaviour>
AUTO | UNINterrupted | MONotone | CVSWr
UNINterrupted|MONotone
Uninterrupted level settings and strictly monotone modes
CWSWr
Constant VSWR
*RST:
AUTO
Example:
SOURce1:POWer:LBEHaviour AUTO
Manual operation:
See
SOURce Subsystem