Remote Control Commands
R&S
®
SMA100B
444
User Manual 1178.3834.02 ─ 03
Parameters:
<Source>
INTernal | EXTernal
*RST:
n.a. (factory preset: INTernal)
Example:
See
Example "Configuring the reference oscillator"
Manual operation:
See
[:SOURce]:ROSCillator:INTernal:TUNing[:STATe]
<State>
Activates the EFC (external frequency control).
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 1)
Example:
See
Example "Configuring the reference oscillator"
Manual operation:
See
[:SOURce]:ROSCillator:INTernal:TUNing:SLOPe
<State>
Sets the sensitivity of the external tuning volatge.
Parameters:
<State>
LOW | HIGH
*RST:
n.a. (factory preset: LOW)
Example:
See
Example "Configuring the reference oscillator"
Manual operation:
See
[:SOURce]:ROSCillator:EXTernal:RFOFf[:STATe]
<State>
Determines that the RF output is turned off when the external reference signal is
selected, but missing.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
n.a. (factory preset: 0)
Example:
See
Example "Configuring the reference oscillator"
Manual operation:
See
"Deactivate RF Output (if external reference is missing)"
[:SOURce]:ROSCillator:EXTernal:FREQuency
<Frequency>
Sets the frequency of the external reference.
Parameters:
<Frequency>
100MHZ | 1GHZ | VARiable | 10MHZ
*RST:
n.a. (factory preset: 10MHZ)