Remote Control Commands
R&S
®
SMA100B
485
User Manual 1178.3834.02 ─ 03
Usage:
Setting only
Manual operation:
See
:TRIGger<hw>:FSWeep[:IMMediate]
:TRIGger<hw>:PSWeep[:IMMediate]
:TRIGger<hw>:LFFSweep:IMMediate
:TRIGger<hw>[:SWEep][:IMMediate]
Performs a single sweep and immediately starts the activated, corresponding sweep:
●
FSWeep
- RF frequency
●
PSWeep
- RF level
●
LFFSweep
- LF frequency
●
SWEep
- all sweeps
Effective in the following configuration:
●
TRIG:FSW|LFFS|PSW|[:SWE]:SOUR
SING
●
SOUR:SWE:FREQ|POW:MODE
AUTO
or
SOUR:LFO:SWE:[FREQ:]MODE
AUTO
Alternativelly, you can use the
IMMediate
command instead of the respective
SWEep:[FREQ:]|POW:EXECute
command.
Example:
TRIG
Starts all active sweeps.
Usage:
Event
Manual operation:
See
:TRIGger<hw>:LFFSweep
Executes an LF frequency sweep in the following configuration:
●
TRIG:LFFS:SOUR
SING
●
LFO:SWE:MODE
AUTO
Example:
LFO:SWE:MODE AUTO
TRIG:LFFS:SOUR SING
TRIG:LFFSweep
Usage:
Event
12.20 UNIT Subsystem
The
UNIT
subsystem is used to set default units for parameters if no unit is indicated in
a command. These settings are valid for the entire instrument.
UNIT Subsystem