Maintenance
R&S
®
SMA100B
495
User Manual 1178.3834.02 ─ 03
In this dialog, you can perform internal calibration routines, and get information on
the last performed calibration.
The remote commands required to define these settings are described in
The information fields show by their background color whether the calibration state of
the instrument is still sufficient. We recommend that you perform internal adjustments
when either the temperature offset or the time since the last full adjustment exceeds
the specified criteria, see
Time Since Last Full Adjustment
and
.
Further functions and individually targeted calibration routines are also available but
password-protected.
●
"Continue Adjustment On Error"
These functions are password-protected. Unlock the protection level 1 to access
them, see
●
Calibration routines
Self-calibration routines that require additional (external) equipment are performed
in the Rohde
&
Schwarz service center.
These functions are password-protected and require higher-level protection pass-
word, see
For more information, see R&S
SMA100B Service Manual.
Adjust All
Performs all available internal calibration routines of the instrument.
Performing Maintenance Tasks