Maintenance
R&S
®
SMA100B
496
User Manual 1178.3834.02 ─ 03
NOTICE!
Risk of DUT damage. During internal adjustments, the instrument temporarily
applies high power at the RF output. This high power can destroy a connected DUT
(device under test).
Do not start internal adjustments if DUT is connected.
Disconnect the DUT and replace it by a terminating resistor with adequate power rat-
ing. We recommend that you use a 50
Ohm, 10
W or larger terminating resistor.
Remote command:
Last Full Adjustment
Displays the date and time of the last fully performed adjustments.
Remote command:
Time Since Last Full Adjustment
Displays the elapsed days since the last full adjustment.
If the last adjustment has been performed more than 100 days ago, the background
color of the parameter turns red.
Temperature Offset To Last Full Adjustment
Displays the temperature difference compared to the calibration temperature.
If the temperature deviates more than
±
5 K, the background of the parameter turns
red.
Remote command:
Information
Displays information to the current adjustment state.
Continue Adjustment on Error
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
This function is password-protected. Unlock the protection level 1 to access it, see
Remote command:
:CALibration<hw>:CONTinueonerror
13.3.4 FPGA/uC Update Settings
Access:
► Select "System Config > Setup > Maintenance > FPGA/µC Update".
Performing Maintenance Tasks