Analog Modulations
R&S
®
SMA100B
97
User Manual 1178.3834.02 ─ 03
The internal LF signal can be used as modulation signal source for any of the ana-
log modulations. The LF signal applies to all modulations which use the internal
modulation signal. Therefore, any modification of the LF signal influences immedi-
ately all currently active modulations.
Settings
.......................................................................................... 97
State (LF frequency sweep)
Activates the generation of an LF frequency sweep signal in the "LF Frequency" dialog.
In the "Sources" tab of the analog modulations, the instrument shows the current state
of the LF frequency sweep.
Note:
Active sweep mode deactivates other sweeps or lists and vice versa.
Remote command:
[:SOURce<hw>]:LFOutput:FREQuency:MODE
Shape
Selects the waveform shape of the LF signal.
The "Triangle" and "Trapezoid" waveforms require option R&S
SMAB-K24.
"Sine"
1 = Period