Varying the RF Signal in List or Sweep Mode
R&S
®
SMA100B
125
User Manual 1178.3834.02 ─ 03
In the "Sources" tab of the analog modulations, the instrument shows the current state
of the LF frequency sweep.
Note:
Active sweep mode deactivates other sweeps or lists and vice versa.
Remote command:
[:SOURce<hw>]:LFOutput:FREQuency:MODE
Current Frequency
In "RF/LF Frequency Sweep" mode, displays the current frequency.
In
= "Step", the parameter is editable and you can enter the next frequency step.
Remote command:
[:SOURce<hw>]:FREQuency:MANual
[:SOURce<hw>]:LFOutput:FREQuency:MANual
Current Level
Applies to "RF Level Sweep" mode.
Displays the current level value.
In
= "Step", the parameter is editable and you can enter the next level step.
Remote command:
Mode
Selects the sweep mode.
For detailed information on the sweep modes and the triggering, see
nal Generation and Triggering in the Sweep and List Modes"
"Auto"
Generates a continuously repeating sweep signal directly after acti-
vating the sweep mode.
The sweep steps are performed automatically, controlled by the dwell
time.
"Single / Extern Single"
Generates a single sweep cycle after a trigger event.
The sweep steps within the cycle are performed automatically, con-
trolled by the dwell time. If one cycle is completed, the instrument
waits for the next trigger event.
"Step / Extern Step"
Generates the sweep signal step-by-step, manually triggered.
"Extern Start/Stop"
Generates a continuously repeating sweep signal that is started,
stopped, and restarted by subsequent external trigger events.
The sweep steps are performed automatically, controlled by the dwell
time.
Remote command:
[:SOURce<hw>]:SWEep[:FREQuency]:MODE
[:SOURce<hw>]:SWEep:POWer:MODE
[:SOURce<hw>]:LFOutput:SWEep[:FREQuency]:MODE
Sweep Mode Settings