File and Data Management
R&S
®
SMA100B
207
User Manual 1178.3834.02 ─ 03
9.9 How to Transfer Files from and to the Instrument
As explained in
on page 190, you access the file system of the
R&S
SMA100B via one of the following ways:
●
Via the built-in "File Manager"
See
Chapter 9.8, "Using the File Manager"
●
On an instrument connected to a LAN:
–
Via one of the standard functions ftp or SMB (samba)
See
Chapter 9.9.2, "Accessing the File System of the R&S
on page 208 and
Chapter 9.9.3, "Accessing the R&S
SMA100B File System via SMB (Samba)"
●
Via a connected USB storage device
See
Chapter 9.9.4, "Using a USB Storage Device for File Transfer"
Mainly because of security reasons, the access to the file system of your
R&S
SMA100B can be denied, because one or all these access methods are deliber-
ately disabled. Access to the file system via LAN and/or USB requires that the corre-
sponding service is enabled and a write access to the file system is enabled. Refer to
Chapter 9.9.1, "Removing File System Protection"
on page 207 for description of the
required steps.
This section provides an introduction to the topic. For comprehensive information, refer
to the application note
1GP72: Connectivity of Rohde&Schwarz Signal Generators
.
Removing File System Protection
.........................................................................207
Accessing the File System of the R&S
......................................208
SMA100B File System via SMB (Samba)
............................. 210
Using a USB Storage Device for File Transfer
......................................................211
9.9.1 Removing File System Protection
Before you try to access the file system via ftp, SMB (samba) or USB, fulfill the follow-
ing:
●
Disable write protection on the file system
●
Enable the corresponding service or interface
To enable write permission on the file system
1. Select "System Config > Setup > Security > Security > General"
2. Enable "Write Nonvolatile Memory"
3. Enter the "Security Password".
The default password is
123456
4. Select "System Config > Setup > Maintenance > Shut Down"
How to Transfer Files from and to the Instrument