Remote Control Commands
R&S
®
SMA100B
308
User Manual 1178.3834.02 ─ 03
*IST?
Individual status query
Returns the contents of the IST flag in decimal form. The IST flag is the status bit
which is sent during a parallel poll.
Return values:
<ISTflag>
0 | 1
Usage:
Query only
*OPC
Operation complete
Sets bit 0 in the event status register when all preceding commands have been execu-
ted. This bit can be used to initiate a service request. The query form writes a "1" into
the output buffer as soon as all preceding commands have been executed. This is
used for command synchronization.
*OPT?
Option identification query
Queries the options included in the instrument. For a list of all available options and
their description refer to the data sheet.
Return values:
<Options>
The query returns a list of options. The options are returned at
fixed positions in a comma-separated string. A zero is returned
for options that are not installed.
Usage:
Query only
Manual operation:
See
*PRE
<Value>
Parallel poll register enable
Sets parallel poll enable register to the indicated value. The query returns the contents
of the parallel poll enable register in decimal form.
Parameters:
<Value>
Range:
0 to 255
*PSC
<Action>
Power on status clear
Common Commands