Remote Control Commands
R&S
®
SMA100B
327
User Manual 1178.3834.02 ─ 03
By default, the instrument uses correction data obtained in the factory before delivery
but customer data can also be used.
You can obtain data for external level correction by using a R&S NRP power sensor.
External level correction is a protected function.
For more information, see R&S
SMA100B Service Manual.
Suffix:
<hw>
.
[1]
Optional suffix
Parameters:
<Data>
FACTory | CUSTomer
*RST:
FACTory
Example:
CALibration:LEVel:EXTern:DATA FACT
// level calibration with data aquired at the factory
CALibration:LEVel:EXTern:EXEC
:CALibration<hw>:LEVel:EXTern:EXEC
Starts level calibaration from the data selected with the command
CALibration<hw>:LEVel:EXTern:DATA
.
Suffix:
<hw>
.
[1]
Optional suffix
Example:
See
:CALibration<hw>:LEVel:EXTern:DATA
Usage:
Event
:CALibration<hw>:LEVel:STATe
<State>
Activates internal level correction.
Suffix:
<hw>
.
[1]
Optional suffix
Parameters:
<State>
OFF | ON | 0 | 1
*RST:
ON
Example:
CALibration:LEVel:STATe 1
:CALibration<hw>:LEVel[:MEASure]?
[<Force>]
Starts level adjustment in the mixer range. The acquired correction values are only
used with FM/Phim or pulse modulation.
CALibration Subsystem