Remote Control Commands
R&S
®
SMA100B
382
User Manual 1178.3834.02 ─ 03
Parameters:
<Source>
SMOothed | FAST
SMOothed
flattens the slew rate, resulting in longer rise/fall times.
FAST
enables fast transitions with shortest rise and fall times.
*RST:
FAST
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:WIDTh
<Width>
Sets the width of the generated pulse, that means the pulse length. It must be at least
20ns less than the set pulse period.
Parameters:
<Width>
float
Range:
20E-9 to 100
Increment: 10E-9
*RST:
2E-6
Example:
See
Example "Perform pulse modulation"
Manual operation:
See
[:SOURce<hw>]:PULM:DOUBle:STATe
<State>
Provided for backward compatibility with former Rohde
&
Schwarz signal generators.
DOUBle
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Pulse Train Settings
Option: R&S
SMAB-K27
This section describes the commands for the pulse train modulation and the associ-
ated file handling. For background information, see
Chapter 4.4.3.2, "Pulse Generator
Example: Generating a pulse train signal
The following settings are required to perform pulse train modulation.
SOURce Subsystem