Remote Control Commands
R&S
®
SMA100B
432
User Manual 1178.3834.02 ─ 03
Parameters:
<Polarity>
NORMal | INVerted
NORMal
Outputs the pulse signal during the pulse width, that means dur-
ing the high state.
INVerted
Inverts the pulse output signal polarity. The pulse output signal
is suppressed during the pulse width, but provided during the
low state.
*RST:
NORMal
Example:
See
Example "Using pulse generator as source for pulse modu-
Manual operation:
See
[:SOURce<hw>]:PGENerator:OUTPut[:STATe]
<State>
Activates the output of the pulse modulation signal.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
See
Example "Using pulse generator as source for pulse modu-
Manual operation:
See
[:SOURce<hw>]:PGENerator:STATe
<State>
Enables the output of the video/sync signal.
If the pulse generator is the current modulation source, activating the pulse modulation
automatically activates the signal output and the pulse generator.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
See
Example "Using pulse generator as source for pulse modu-
12.15.9 SOURce:PHASe Subsystem
This subsystem contains the commands for adjusting the phase of the RF output signal
relative to a reference signal of the same frequency.