Remote Control Commands
R&S
®
SMA100B
445
User Manual 1178.3834.02 ─ 03
Example:
See
Example "Configuring the reference oscillator"
Options:
VARiable requires R&S
SMAB-K704
100MHZ|1GHZ require R&S
SMAB-K703
Manual operation:
See
"External Reference Frequency"
[:SOURce]:ROSCillator:EXTernal:FREQuency:VARiable
<Frequency>
Specifies the user-defined external reference frequency.
Parameters:
<Frequency>
float
Range:
1E6 to 100E6
Increment: 0.1
*RST:
n.a. (factory preset: 1E7)
Default unit: Hz
Example:
See
Example "Configuring the reference oscillator"
Options:
R&S
SMAB-K704
Manual operation:
See
"Variable Reference Frequency"
[:SOURce]:ROSCillator:EXTernal:SBANdwidth
<SBandwidth>
Selects the PLL synchronization bandwidth depending on the used external reference
signal.
For more information, see data sheet.
Parameters:
<SBandwidth>
WIDE | NARRow
NARRow
The synchronization bandwidth is a few Hz.
Suitable for external reference sources with phase noise worse
than the R&S
SMA100B.
WIDE
Uses the widest possible synchronization bandwidth; achieves
best residual phase noise performance and phase stability
between the synchronized devices.
For more information, see data sheet.
*RST:
n.a. (factory preset: WIDE)
Example:
See
Example "Configuring the reference oscillator"
Manual operation:
See
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE
<OutpFreqMode>
Sets the output reference frequency.
SOURce Subsystem