Remote Control Commands
R&S
®
SMA100B
447
User Manual 1178.3834.02 ─ 03
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]
<State>
Determines whether the calibrated (off) or a user-defined (on)
adjustment value
is
used for fine adjustment of the frequency.
Parameters:
<State>
0 | 1 | OFF | ON
0
Fine adjustment with the calibrated frequency value
1
User-defined adjustment value
The instrument is no longer in the calibrated state.
The calibration value is, however, not changed. The instrument
resumes the calibrated state if you send
SOURce:ROSCillator:INTernal:ADJust:STATe 0
.
*RST:
n.a. (factory preset: 0)
Example:
// query
calibration
value
CALibration:ROSCillator?
// 32767
// Set an internal source
// Activate user-defined
adjustment
value of 1000
SOURce:ROSCillator:SOURce INT
SOURce:ROSCillator:INTernal:ADJust:STATe 1
SOURce:ROSCillator:INTernal:ADJust:VALue 1000
// to resume calibrated state
SOURce:ROSCillator:INTernal:ADJust:VALue 0
SOURce:ROSCillator:INTernal:ADJust:STATe 0
// or
// SYSTem:FPRest
Manual operation:
See
12.15.12 SOURce:SWEep Subsystem
The
SOURce:SWEep
subsystem contains the commands for configuring RF sweep sig-
nals.
SOURce Subsystem