Remote Control Commands
R&S
®
SMA100B
474
User Manual 1178.3834.02 ─ 03
Usage:
Query only
Manual operation:
See
:SYSTem:SRData?
Queris the SCPI recording data from the internal file.
This feature enables you to transfer an instrument configuration to other test environ-
ments, as e.g. laboratory virtual instruments.
Return values:
<FileData>
block data
Example:
SYSTem:SRData?
// #3118:SOURce1:ROSCillator:SOURce EXT
:SOURce1:FREQuency:CW 4000000000
:SOURce1:FREQuency:OFFSet 1000000
:SOURce1:AM1:STATe 1
:OUTPut1:STATe 1
Usage:
Query only
:SYSTem:STARtup:COMPlete?
Queries if the startup of the instrument is completed.
Return values:
<Complete>
0 | 1 | OFF | ON
*RST:
0
Example:
SYST:STAR:COMP?
Response:
1
the startup of the instrument is completed.
Usage:
Query only
:SYSTem:DATE
<Year>, <Month>, <Day>
Queries or sets the date for the instrument-internal calendar.
This is a password-protected function. Unlock the protection level 1 to access it, see
.
Parameters:
<Year>
integer
<Month>
integer
Range:
1 to 12
<Day>
integer
Range:
1 to 31
SYSTem Subsystem