Troubleshooting and Error Messages
R&S
®
SMA100B
502
User Manual 1178.3834.02 ─ 03
The message is displayed until the error is eliminated. It covers the status display in
the info line. After error elimination, the message automatically disappears and is also
recorded in the history.
14.2 SCPI-Error Messages
The SCPI error messages are the same in all SCPI instruments. Detailed information
and an overview of all error messages as defined in SCPI standard can be found in the
corresponding documentation.
The errors are assigned negative numbers. The error text being entered into the error/
event queue or being displayed is printed in bold face on the left together with the error
code. Below the error text, there is an explanation as to the respective error.
14.3 Device-Specific Error Messages
The following table contains all error messages specific for the instrument in alphabeti-
cal order, as well as an explanation of the error situation. The positive error codes
mark the errors specific to the instrument.
The device-specific error messages set bit 3 in the ESR register.
The index provides a list of the error messages sorted according to their error codes.
Error
Code
Error
Description
Remedy
50
Extern reference out of
range or disconnected
External reference is selected but no external sig-
nal is applied or the signal is out of range.
●
Check the selected reference signal
source (internal or external) in the
"Setup > Reference Oscillator" dia-
log.
●
Change setting to 'internal' if no
appropriate external source is avail-
able.
140
This modulation forces
other modulations off
A modulation has been switched on which cannot
be used at the same time as an already active
modulation. The previous modulation has been
switched off.
180
Adjustment failed
Adjustment could not be executed
The adjustment data have to be gener-
ated first by an internal or external
adjustment or to be loaded into the
device
182
Adjustment data missing
Adjustment data is missing.
The adjustment data has to be generated
first by an internal or external adjustment
or to be loaded into the instrument.
Device-Specific Error Messages