Instrument Tour
R&S
®
SZU100A
17
Manual 1425.3061.02 ─ 04
LED
R&S
SZU status
Description
on
Internal error
E.g. over temperature, power failure, fan failure.
blinking (6Hz)
External error
E.g. LO level out of range.
6.2.4
USB Interface
Universal serial bus (USB) interface of type B (
to communicate with the host instrument, the R&S
SMW.
6.2.5
LO In/Out Connectors
Connectors for RF signal input and output (
, (5)). These connectors are
SMA coaxial connectors with 50
Ohm impedance. LO In receives the LO signal
from the R&S
SMW, and LO Out is intended for future use, e.g. to forward the LO
signal when using multiple R&S
SZU in a daisy chain.
Risk of instrument damage
The RF input power at the connector LO In must not exceed the maximum
values specified in the data sheet.
6.2.6
Trigger Input/Output Connector
Connector for an optional trigger signal (
, (6)). The trigger connector is
an SMA coaxial connector with 50
Ohm impedance.
Rear View