Troubleshooting
R&S
®
TSMA6
94
User Manual 4900.8057.02 ─ 06
8.2
Evaluate Selftest File
●
Path:
C:\ProgramData\Rohde-Schwarz\TSMA\Selftestresults
●
Filename:
Selftestresult.txt
Entries in the
Selftestresult.txt
are created during the boot process of the
device and every 5 minutes.
To open the
Selftestresult.txt
, proceed as follows.
1. Open a file explorer.
2. Enter the path to the directory where the selftest file is stored.
3. Open the selftest file.
Following device parameters and values are stored.
●
Internal voltages (nominal values / actual values)
●
Temperatures (scanner / mainboard)
●
Battery status info
●
Status of fans
●
Status of scanner interface
8.3
Verify Scanner Link / Recall Device Info Using Tsme-
DeviceManager
1. To start the R&S TSME Device Manager, double-click the desktop icon or select in
Windows:
"Start" > "All Programs" > "TsmeTools" > "TsmeDeviceManager"
2. Following error messages may occur.
●
Tsme Device Manager Application has stopped
In the web GUI. check "Mode of Operation". If necessary, change to "PC Mode"
and try again.
●
No TSME connected
In the web GUI, verify "Scanner LAN Port Settings".
3. If the"Device Info" tab (see
) contains scanner data, the scanner connec-
tion has been established successfully.
4. Check the "Device Analysis Output" (see
Chapter 8.3.2, "Device Analysis Output"
on page 96) and the "Options" (see
Chapter 8.3.3, "Verify Installed License
Verify Scanner Link / Recall Device Info Using TsmeDeviceManager