The calibration type depends on the number of ports and of the analyzer type. If a
single port is calibrated, the analyzer uses a reflection calibration type (e.g. Full
One Port / Refl OSM). For 2 and more ports, you can choose among several cali-
bration types.

4.4

 

User characterization

The calibration unit offers the possibility to store multiple user characterization
data files either on its internal flash memory or on the microSD card (R&S ZN-
Z152, R&S ZN-Z153). Use the "Characterize Cal Unit" function of your network
analyzer for this purpose. For details, refer to the analyzer’s help system.

4.5

 

microSD card

On the rear of the R&S ZN-Z152 and R&S ZN-Z153 calibration units, there is a
microSD card slot. The microSD card can be used to store user characterization
data. The factory data is always stored on the internal memory.

Use the microSD card shipped with the calibration unit, if possible.

Before removing or inserting the microSD card, disconnect the USB
connection between calibration unit and network analyzer.

4.6

 

Mounting the calibration unit

To mount the calibration unit e.g. on a wafer prober, remove the elastic buffers on
the bottom of the housing. There are either two threads (R&S ZN-Z156) or four
threads for use of M2.5 screws.

Mounting the calibration unit

Summary of Contents for ZN-Z15 Series

Page 1: ...R S ZN Z15x Calibration Units User Manual 1337596302 Version 04 U 2...

Page 2: ...f 2 ports 5 GHz to 67 GHz 1332 7239 02 R S ZN Z156 1 85 mm f 2 ports 10 MHz to 67 GHz 1332 7239 03 User manual 1326 5080 02 covers the R S ZN Z154 calibration unit 2022 Rohde Schwarz GmbH Co KG Muehl...

Page 3: ...roduction and verification of electronic components and devices in industrial administrative and laboratory environments Use the R S ZN Z15x only for its designated purpose Observe the operating condi...

Page 4: ...ning messages in the documentation A warning message points out a risk or danger that you need to be aware of The signal word indicates the severity of the safety hazard and how likely it will occur i...

Page 5: ...do not wear out because they are switched electroni cally The R S ZN Z15x calibration units differ in the number of calibration ports and the connector types For an overview refer to the data sheet 3...

Page 6: ...esidential buildings Class A equipment is intended for use in industrial environments It can cause radio disturbances in residential environments due to possible conducted and radiated disturbances It...

Page 7: ...nts of the product and the device under test DUT Ground yourself to prevent electrostatic discharge damage a Use a wrist strap and cord to connect yourself to ground b Use a conductive floor mat and h...

Page 8: ...See also Cable selection and electromagnetic interference EMI on page 6 3 5 RF connections The calibration unit provides two RF connectors numbered 1 to 2 to be connec ted to the test ports of the ana...

Page 9: ...ialization the status LED will switch to green see Chapter 4 2 Function of the status LED on page 9 The unit is ready to use as outlined below 4 2 Function of the status LED The LED on top of the Cali...

Page 10: ...ssignment a Connect the calibration unit ports to the test ports of the network analyzer as depicted in the analyzer GUI b Start the calibration data acquisition 4 Disconnect the RF ports of the calib...

Page 11: ...of your network analyzer for this purpose For details refer to the analyzer s help system 4 5 microSD card On the rear of the R S ZN Z152 and R S ZN Z153 calibration units there is a microSD card slot...

Page 12: ...ct contact our customer support center A team of highly qualified engineers provides support and works with you to find a solution to your query on any aspect of the operation program ming or applicat...

Reviews: