Preparing for use

R&S

®

ZN-Z15x

5

User Manual 1337.5963.02 ─ 04

Automatic calibration is faster and less error-prone than manual calibration:

There is no need to connect calibration standards manually, which is particu-
larly time-saving if you want to calibrate multiple ports.

Invalid calibrations due to operator errors (e.g. wrong standards or improper
connections) are almost excluded.

There is no need to handle calibration kit data.

The internal standards do not wear out because they are switched electroni-
cally.

The R&S

 

ZN-Z15x calibration units differ in the number of calibration ports and

the connector types. For an overview, refer to the data sheet.

3

 

Preparing for use

Here, you can find basic information about setting up the product for the first time.

3.1

 

Unpacking and checking

1. Unpack the product carefully.

2. Retain the original packing material. Use it when transporting or shipping the

product later.

3. Using the delivery notes, check the equipment for completeness.

4. Check the equipment for damage.

If the delivery is incomplete or equipment is damaged, contact
Rohde

 

&

 

Schwarz.

Unpacking and checking

Summary of Contents for ZN-Z15 Series

Page 1: ...R S ZN Z15x Calibration Units User Manual 1337596302 Version 04 U 2...

Page 2: ...f 2 ports 5 GHz to 67 GHz 1332 7239 02 R S ZN Z156 1 85 mm f 2 ports 10 MHz to 67 GHz 1332 7239 03 User manual 1326 5080 02 covers the R S ZN Z154 calibration unit 2022 Rohde Schwarz GmbH Co KG Muehl...

Page 3: ...roduction and verification of electronic components and devices in industrial administrative and laboratory environments Use the R S ZN Z15x only for its designated purpose Observe the operating condi...

Page 4: ...ning messages in the documentation A warning message points out a risk or danger that you need to be aware of The signal word indicates the severity of the safety hazard and how likely it will occur i...

Page 5: ...do not wear out because they are switched electroni cally The R S ZN Z15x calibration units differ in the number of calibration ports and the connector types For an overview refer to the data sheet 3...

Page 6: ...esidential buildings Class A equipment is intended for use in industrial environments It can cause radio disturbances in residential environments due to possible conducted and radiated disturbances It...

Page 7: ...nts of the product and the device under test DUT Ground yourself to prevent electrostatic discharge damage a Use a wrist strap and cord to connect yourself to ground b Use a conductive floor mat and h...

Page 8: ...See also Cable selection and electromagnetic interference EMI on page 6 3 5 RF connections The calibration unit provides two RF connectors numbered 1 to 2 to be connec ted to the test ports of the ana...

Page 9: ...ialization the status LED will switch to green see Chapter 4 2 Function of the status LED on page 9 The unit is ready to use as outlined below 4 2 Function of the status LED The LED on top of the Cali...

Page 10: ...ssignment a Connect the calibration unit ports to the test ports of the network analyzer as depicted in the analyzer GUI b Start the calibration data acquisition 4 Disconnect the RF ports of the calib...

Page 11: ...of your network analyzer for this purpose For details refer to the analyzer s help system 4 5 microSD card On the rear of the R S ZN Z152 and R S ZN Z153 calibration units there is a microSD card slot...

Page 12: ...ct contact our customer support center A team of highly qualified engineers provides support and works with you to find a solution to your query on any aspect of the operation program ming or applicat...

Reviews: