Command reference
R&S
®
ZNB/ZNBT
1036
User Manual 1173.9163.02 ─ 62
Example:
CALC:TRAN:VNET:GLO:EMB:TND SL
Select the Shunt L circuit model for embedding.
CALC:TRAN:VNET:GLO:EMB:PAR:R SL, 2.2E+3; :CALC:
TRAN:VNET:GLO:EMB ON
Increase the resistance for the Shunt L circuit model to 2.2 kΩ
and enable embedding.
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:
TNDefinition
<CircuitModel>
Selects the circuit model for ground loop embedding.
Suffix:
<Ch>
.
Channel number.
<group>
Port group (DUT) number.
If multiple port groups are configured (see
ON
, then each port group can
have its own de-/embedding models.
Parameters:
<CircuitModel>
FIMPort | SL | SC | SG
Possible circuit models (character data); see
.
*RST:
FIMPort
Example:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:
Manual operation:
See
CALCulate<Ch>:TRANsform:VNETworks:GLOop:GROup
<Boolean>
If set to ON (1), each port group (defined using
)
has its own ground loop embedding and deembedding network. Otherwise the same
embedding network and deembedding network is used for all active ports.
See the
CALCulate<Ch>:TRANsform:VNETworks:GLOop:DEEMbedding<group>:...
and
CALCulate<Ch>:TRANsform:VNETworks:GLOop:EMBedding<group>:...
commands.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
Range:
ON (1) | OFF (0)
*RST:
OFF (0)
Manual operation:
See
SCPI command reference