Command reference
R&S
®
ZNB/ZNBT
1180
User Manual 1173.9163.02 ─ 62
●
The signals are switched on as soon as a measurement in a segment with non-
zero segment bits is started. They are changed whenever a segment with different
segment bits is measured.
●
The signals at the USER PORT connector are maintained after the analyzer enters
the hold state. This happens in single sweep mode after all sweep sequences have
been terminated.
Tip:
You can use the active segment number as a parameter for
OUTPut<Ch>:UPORt:SEGMent<Seg>[:VALue]
and monitor the measurement in up
to 16 different segments per channel at the USER PORT connector; see example
below. You can also use the USER PORT output signals as segment-dependent trigger
signals for external devices. Use
CONTrol:AUXiliary:C[:DATA]
to transfer the
four bit value in decimal representation.
Suffix:
<Ch>
.
Channel number.
<Seg>
Sweep segment number
Parameters:
<BinValue>
Binary value. The transferred values correspond to the following
states of the USER CONTROL connector:
#B0000 - no signal at any of the four pins 16, 17, 18, 19
#B0001 - output signal at pin 16
#B0010 - output signal at pin 17
#B0011 - output signal at pin 16 and 17
...
#B1111 - output signal at pin 16, 17, 18 and 19
Range:
#B0000 to #B1111 (for setting command), 0 to 15
(query)
*RST:
#B0000 (0)
SCPI command reference