Command reference
R&S
®
ZNB/ZNBT
1210
User Manual 1173.9163.02 ─ 62
A complete, valid set of port assignments must be defined before you can initiate a cal-
ibration. See
Chapter 4.5.5.6, "Multiple port assignments"
information.
Suffix:
<Ch>
.
Number of the calibrated channel
<Asg>
Number of the port assignment
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
Manual operation:
See
"Start Cal Sweep / Abort Sweep"
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:ALL:COUNt?
Returns the total number of port assignments of
all
calibrations.
See
Chapter 4.5.5.6, "Multiple port assignments"
on page 191 for background infor-
mation.
Note
:
Use
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
to get the
number of port assignments of the "current" calibration.
Suffix:
<Ch>
.
Channel number
Example:
See
Chapter 8.2.5.2, "MultiCal (with calibration unit)"
Usage:
Query only
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
Returns the number of port assignments of the "current" calibration, i.e. the calibration
last created using
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
This is particulary useful if the default port assignments were established using
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
.
See
Chapter 4.5.5.6, "Multiple port assignments"
on page 191 for background infor-
mation.
Note
:
In MultiCal scenarios, use
[SENSe<Ch>:]CORRection:COLLect:AUTO:
to get the number of port assignments of all calibrations.
Suffix:
<Ch>
.
Channel number
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
SCPI command reference