Command reference
R&S
®
ZNB/ZNBT
1224
User Manual 1173.9163.02 ─ 62
Example:
See
Chapter 8.2.5.5, "Inline calibration"
Usage:
Event
[SENSe<Ch>:]CORRection:COLLect:AUTO:SAVE
Calculates the system error correction for automatic calibrations with
, saves the data and applies the calibration to the active channel.
Requires successful calibration sweeps for all related port assignments (see
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment<Asg>:ACQuire
)
Suffix:
<Ch>
.
Channel number of the calibrated channel
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:TEMPcomp
<Boolean>
Available if the calibration unit selected using
(ICS) only. Applies to all inline calibra-
tion units (ICUs) that are connected to the selected ICS and that are used with factory
characterization data .
If set to ON (1), the R&S
ZNB/ZNBT firmware adjusts the factory characterization data
according to the ICU temperatures.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean>
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
<CalType>[, <Characterization>[,
<TestPort1>[, <TestPort2>]...]]
Selects and initiates an automatic calibration at arbitrary analyzer and cal unit ports.
This command also selects the calibration type. A progress monitor for the calibration
sweeps is displayed.
Tip:
If the test setup contains a high attenuation the analyzer may fail to detect the cal
unit ports connected to each of its ports. In this case use the extended command
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
.
If several calibration units are connected, use
to select a unit for the calibration.
SCPI command reference