Command reference
R&S
®
ZNB/ZNBT
1337
User Manual 1173.9163.02 ─ 62
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected frequency applies to all source ports in the active
channel.
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency.
*RST:
1 GHz
Default unit: Hz
Example:
FUNC "XTIMe:POW:A1"
Activate a time sweep and select the wave quantity a
1
as mea-
sured parameter for channel and trace no. 1.
FREQ:CW 100MHz
Set the CW frequency to 100 MHz.
Manual operation:
See
7.3.15.2
SOURce:GROup...
The
SOURce:GROup...
commands define the group of (un-)used logical ports.
........................................................................................ 1337
.............................................................................. 1338
.....................................................................................1338
SOURce<Ch>:GROup<Grp>:DPORt:COUNt
.................................................................. 1339
.............................................................................. 1339
........................................................................... 1339
............................................................................. 1340
SOURce<Ch>:GROup<Grp>:PPORt<PhyPort>:DPORt
................................................... 1340
............................................................................1341
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:CONDition
......................................1342
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:BWFactor
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:DVALue
........................... 1344
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:MOFFset:MODE
............................. 1345
SOURce<Ch>:GROup<Grp>:SIMultaneous:FOFFset:STATe
............................................ 1345
SOURce<Ch>:GROup<Grp>
<arg0>, <arg1>
In "set direction" this command defines the set of active logical ports (the "master
group") as the continuous range from
<log_port1>
to
<log_port2>
, i.e. logical
ports outside this range are disabled.
In "get direction" it returns the minimum and maximum enabled logical port for the rela-
ted group.
See
on how to define logical ports.
Note:
Setting
SOURce<Ch>:GROup
or
dissolves the port groups created using
.
SCPI command reference