Command reference
R&S
®
ZNB/ZNBT
1434
User Manual 1173.9163.02 ─ 62
Example:
SEGM:ADD; :SWE:TYPE SEGM
Select segmented frequency sweep.
TRIG:LINK 'SEGM'
Select a trigger segment as triggered measurement sequence.
TRIG:LINK?
Query the triggered measurement sequence. The response is
'SEGMENT'
.
Manual operation:
See
TRIGger<Ch>[:SEQuence]:MULTiple:HOLDoff
<MeasSequence>[, <DelayTime>]
Sets a delay time between the trigger event and the start of the measurement ("Trigger
Delay") in multiple trigger mode.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasSequence>
SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence,
PPOint
denotes "partial
measurement"; see
<DelayTime>
Delay time
Range:
0 s to 13680 s
Increment: 10 ms
*RST:
0 s
Default unit: s
Example:
See
TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
Manual operation:
See
TRIGger<Ch>[:SEQuence]:MULTiple:SLOPe<Num>
<MeasSequence>[, <Slope>]
Qualifies whether the multiple trigger events occur on the rising or on the falling edge
or on the beginning of the high / low level periods of the external TTL trigger signal
<Num>
.
Suffix:
<Ch>
.
Channel number
<Num>
Number of trigger signal (1 or 2)
Parameters:
<MeasSequence>
SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence,
PPOint
denotes "partial
measurement"; see
<Slope>
POSitive | NEGative | HIGH | LOW
Trigger slope for the triggered measurement sequence:
POSitive | NEGative - rising or falling edge
HIGH | LOW - high or low level
SCPI command reference