GUI reference
R&S
®
ZNB/ZNBT
308
User Manual 1173.9163.02 ─ 62
5.2.7
Y-Z-Params tab
Allows you to select Y- and Z-parameters as measured quantities. Both Y- and Z-
parameters can serve as an alternative to S-parameters for characterizing a linear n-
port network.
Background information
Refer to the following sections:
●
Chapter 4.3.3, "Impedance parameters"
●
Chapter 4.3.4, "Admittance parameters"
●
Chapter 4.3.3.2, "Z-parameters"
●
Chapter 4.3.4.2, "Y-parameters"
Y/Z-Parameter
or
as a measured quantity for the active trace.
For an n-port vector network analyzer, the pull-down list provides the full set of n
2
Y-
and Z-parameters.
Y- and Z-parameters are expressed as Y/Z
<out>< in>
, where <out> and <in> denote the
output and input port numbers of the DUT.
Remote command:
CALCulate<Ch>:PARameter:MEASure
"<Trace_Name>", "Y11" |
"Z11" ...
CALCulate<Ch>:PARameter:SDEFine
"<Trace_Name>", "Y11" |
"Z11" ...
Y11/Y12 /Y21/Y22
Selects one of the 2-port
Chapter 4.3.4.2, "Y-parameters"
on page 134 as a measured
quantity for the active trace. The Y-parameters describe the admittances of a DUT with
output ports terminated in a short circuit (V = 0).
Meas softtool